❄️Cooling down with LHe

Fig. 1. Cooling system of 1-Squared in action, overview
  1. Check that helium line is deep inside the dewar, ca. 5 cm above the bottom.

  2. Check that Helium pump is running (see panel inside the frame from the side of the load lock):

Helium pump button (red arrow)

3. Check that black valve below the frame is open:

In this configuration gas is passing through a black valve reducing the flow. To fully open rotate 180 deg.

4. Make sure that helium recovery line is connected to a He return port on the wall-panel.

5. Use ladder and open the needle valve 1 full turn. Rotate counter clockwise if you look from below.

Opening direction

6. Wait until temperature reading starts to steadily go down by 0.01 K. If this doesn't happen in 10 minutes - it is likely that there is a problem (*).

7. When sample temperature reaches ca. 40K start closing needle valve in small steps and wait for the temperature to stabilize. Try to close as much as possible to reduce helium consumption, but don't let temperature to go up.

8. The lowest sample temperature reading is currently ~26 K (condition on Jan 2025).

To reach the lowest possible temperature on the sample one can improve thermal contact by tightening the screws on the sample pocket of manipulator head. See here for more details.

Dynamics of manipulator cooling

T vs time for the first cooling down and for setting temperatures using the manipulator heater.

Troubleshooting:

(*) As a first step try to disconnect helium recovery line from the wall (see Fig. 1) and check if there is a flow of gas coming out of it.

Last updated

Was this helpful?