Starting the experiment
Last updated
Was this helpful?
Last updated
Was this helpful?
Open the valve between Preparation and Analysis chambers;
in the Analysis chamber;
Make sure that beam is not used and beamshutter is closed;
Take the beam using ;
Switch on the ;
Set , , [width of apertures (optional)];
Check that suitable and are being used;
(data types, folder, sample name, etc.);
Switch on analyzer electronics by tumbler and open the beamline valves (starting from Experiment);
Open in the SES menu "Calibration" -> "Voltages" and set kinetic energy where you expect to see the bands;
Find signal from the sample using joystick and camera images. Sticker roughly marks position of the beam (camera image under the sticker should be in focus).
Fine tune the position using "Manual/Steps" or "Virtual axes" tabs. Y direction is along the beam (use for focusing).
Have a lot of nice measurements!