  1. User manuals 1^2

Experts menu

Help for beamline scientists in troubleshooting

PreviousOperating wobble stickNextSputtering

Last updated 1 year ago

Was this helpful?

After manipulator went crazy

! Press red button immediately !

  1. Close connection in MultiAx (tab Settings)

  2. Manipulator -> OFF (button on the back panel, next to LL vacuum gauge)

  3. Release the red button

  4. Wait 30 seconds

  5. Manipulator -> ON

  6. Watchdog -> OFF. You will need admin password for this.

  7. Open connection in MultiAx

  8. Watchdog -> ON.

Recalibrate axis