T-dependent measurements
Using manipulator heater to set sample temperature
Last updated
Using manipulator heater to set sample temperature
Last updated
Make sure that Contol A LED is on (the coldfinger temperature will be used as a reference).
On the Lakeshore TC press Setpoint button (6).
Dial in the desired temperature for coldfinger (A)*. Sample temperature will be ~10 deg higher.
Press Enter.
Press Heater Range button and select power. Start with low, if it is not enough - try medium.
Press Enter.
Find below an approximate table of correspondence between A and B temperatures.
10
19.5
low
18
25
?
24
30
med
29
35.2
med
33.1
40
med
37
45
med
41
50
med
45.3
55.2?
med
49
60
med
57
70.5
high
85
100
high